2a - v1.0 FACTORY CALIBRATION

Overview

Description

Do not recalibrate you device unless you know what you're doing! If you have questions or concerns about your device, please ask on the forums

This calibration includes:

  1. Calibrating lights 1 - 10 to PAR against two LiCOR sensors attached to the MultispeQ by using a stock MultispeQ as a known light source

Directions

Once at the beginning of the day, make sure to do the following steps before calibrating the LED’s on new devices

  1. Ensure that the project 0 - v1.0 STOCK MULTISPEQ CALIBRATION is completed.
  2. Connect to the Stock device in the Desktop App (microUSB to USB). In the console which can be accessed by clicking the gear icon labeled "settings". Enter command "calibrate_leds+". The device will then run the light sequence for calibrating LEDs when prompted via the USB-C cable. Once you have entered the command, you can disconnect the stock device from the Desktop App. Unless you are using Windows computers, than DO NOT DISCONNECT.

Once that is complete, you can calibrate many devices using the following directions:

  1. Connect the black USB-C cable to the Stock device and the white USB-C cable to the device to be calibrated.
  2. Connect the device to be calibrated to the Desktop App.
  3. Place the cabled LiCORs on the stock device. Begin collecting data. Lights on the stock device should flash.
  4. Leave the cables hooked up in this configuration to complete the next calibration project: 2b – v1.0 FACTORY CALIBRATION

Once complete, evaluate the quality of the results for each measurement (there are several so scroll down to see them all. If there are danger or warning messages read and retry the calibration. Then...

  1. Hit the "save to Device" button. Then submit the data to the website.

Protocols

v1.0 CALIBRATION: NEW LED PAR a

Description

Used to calibrate the Light Calibration Box to a LiCOR PAR. Used only in factory calibration.

Do not run this on the device you are calibrating! Instead, run this on the raw PCB which is supplying the light for the calibrating board.

v1.0 CALIBRATION: NEW LED PAR b

Description

Used to calibrate the Light Calibration Box to a LiCOR PAR. Used only in factory calibration.

Do not run this on the device you are calibrating! Instead, run this on the raw PCB which is supplying the light for the calibrating board.

v1.0 CALIBRATION: NEW LED PAR c

Description

Used to calibrate the Light Calibration Box to a LiCOR PAR. Used only in factory calibration.

Do not run this on the device you are calibrating! Instead, run this on the raw PCB which is supplying the light for the calibrating board.

v1.0 CALIBRATION: NEW LED PAR d

Description

Used to calibrate the Light Calibration Box to a LiCOR PAR. Used only in factory calibration.

Do not run this on the device you are calibrating! Instead, run this on the raw PCB which is supplying the light for the calibrating board.

Questions

  1. Device ID (Short Answer)

Locations

Team

Thumb screen shot 2017 05 31 at 12.23.10 pm
Project Lead (0)
Thumb 012409 pbi 160 low low res headshot
Collaborator (34)
Default avatar
Collaborator (31)
Default avatar
Collaborator (387)
Default avatar
Collaborator (375)
Default avatar
Collaborator (167)
Thumb sam 0645   copy   copy  2
Collaborator (0)