calibrate_external_RGBW_using_LiCOR_sensor


This is only to be used with the PhotosynQ calibration system!

Sends a series of red, green, blue and white light intensities using a "neopixel" type RGBW LED connected to the PAR calibration system.

Collects light intensities using a PAR sensor, which is also connected to the PAR calibration system.

Finds a poly-linear relationship that allows the prediction of PAR from the LED input parameters.

This calibrated system can then be used to calibrate other PAR devices.

[
  {
    "set_repeats": 12,
    "v_arrays": [
      [
        20,
        40,
        80,
        100,
        120,
        140,
        160,
        180,
        200,
        220,
        240,
        0
      ]
    ],
    "v_idx": "s",
    "_protocol_set_": [
      {
        "indicator": [
          "@0",
          0,
          0,
          0
        ],
        "ext_par": [
          100
        ]
      },
      {
        "indicator": [
          0,
          "@0",
          0,
          0
        ],
        "ext_par": [
          100
        ]
      },
      {
        "indicator": [
          0,
          0,
          "@0",
          0
        ],
        "ext_par": [
          100
        ]
      },
      {
        "indicator": [
          0,
          0,
          0,
          "@0"
        ],
        "ext_par": [
          100
        ]
      },
      {
        "indicator": [
          "@0",
          "@0",
          "@0",
          "@0"
        ],
        "ext_par": [
          100
        ]
      }
    ]
  }
]
Dave in cap img 4615
Created by

David M. Kramer


Category

Miscellaneous

Compatible Instruments

MultispeQ v1.0


Latest Update

May 2018