_protocol_set_

Treating protocols as integrated sets.

Protocols are separate units which produce separate data sets. Using _protocol_set_, multiple protocols can be chained together into one protocol.

Note: This is different from chaining protocols together in a project. There, the protocols are still separate units within a measurement.

Input: array

Example:

"_protocol_set_": [{ Protocol 1 },...],

Editor: fixed

Instruments:

MultispeQ 1


adc_show deprecated

When "adc_show" is 1, the readings taken by the ADC (analog to digital converter) on the sample and hold circuit are outputted in "data_raw" instead of the normal output. No other output is recorded, and only the last set of adc readings is shown (all other readings are not outputted). By default there are 19 samples, but this can be changed using "number_samples".

Input: number

Values:

  • 0, 1

Example:

"adc_show": <number>,

Editor: fixed

Instruments:

MultispeQ 1


autogain

The autogain function is added to the main section of the protocol. Multiple autogain functions can be run in a single protocol, and the indexed values should be useable throughout the experiment.

<index> which index to store gain settings (0-9)
<pulsed_LED> which pulsed LED to test (0-9)
<detector> which detector to use (0-3)
<pulse_duration> what duration to use (in microseconds) (1-200)
<target_value> what target value to use (0-65535, recommended: 40000-50000)

Input: nested array

Example:

"autogain": [ [<index>, <pulsed_LED>, <detector>, <pulse_duration>, <target_value>],... ],

Editor: fixed

Instruments:

MultispeQ 1


averages

averages sets the number of times to average the protocol. The protocol will be repeated equal to the number of averages, and the resulting data will be averaged and outputted as a single data point. Averaging is often used to reduce noise and improve the quality of measurements.

Input: number

Values:

  • 0 - 10000

Example:

"averages": <number>,

Editor: fixed

Instruments:

MultispeQ 1


averages_delay

"averages_delay" defines the delay between protocol averages in in milliseconds (ms).

Input: auto

Values:

  • 0 - 9999999999

Example:

"averages_delay": <number>,

Editor: fixed

Instruments:

MultispeQ 1

Dependancies:

  • averages

dac_lights

When "dac_lights" is set to 1, "pulsed_lights_brightness" and "nonpulsed_lights_brightness" settings will be interpreted by the device as 12-bit values (0 - 4095) directly controlling the LED voltage (via the DAC), instead of microEinsteins (uE). This is used during factory calibration.

Note: Setting brightness to high using raw DAC values could permanently disable LEDs, so be careful!

Input: number

Values:

  • 0, 1

Example:

"dac_lights": <number>,

Editor: fixed

Instruments:

MultispeQ 1


detectors

"detectors" defines which detectors are being measured, and in what order. The chosen "meas_light" supplies the light source to be measured. The detector measurement is recorded in "data_raw" in the data received from the MultispeQ, which is then graphed for the user. When "detectors" is set to 0, no detector is read and responses will be recorded as zero.

Input: nested array

Values:

  • 0 - 4

Example:

"detectors": [[<detector>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_length,+ pulse_distance,+ pulsed_lights,+ pulsed_lights_brightness

energy_min_wake_time deprecated

The value of time is in milliseconds, and must be between zero and 10^6 milliseconds. This setting changes the delay between shutting off the 5V (following energy_save_timeout) and the wake up time. The energy_min_wake_time is needed to prevent brown out following the initiation of energy save.

Important: Note that the change in energy_min_wake_time will remain in effect until the instrument is reset, when it will return to the default value of 10 s.

Input: number

Values:

  • 0 - 10^6

Example:

"energy_min_wake_time": <number>,

Editor: fixed

Instruments:

MultispeQ 1


energy_save_timeout

Adjusting Energy Save Timeout Time. The value of time is in milliseconds, and must be between zero and 10^6 milliseconds. This setting changes the energy_save_timeout time to prevent interference from the requirement for a 7 second wake up time from power save mode.

Important: Note that the change in energy_save_timeout will remain in effect until the instrument is reset, when it will return to the default value of 120 s.

Input: number

Values:

  • 0 - 10^6

Example:

"energy_save_timeout": <number>,

Editor: fixed

Instruments:

MultispeQ 1


environmental

environmental defines which additional sensor(s) to measure. The environmental calls occur at the very beginning of the measurement, prior to any pulses. Some sensors require additional information, like specifying which pin to measure, brightness, etc. In addition to sensor measurements, there are also calls to flip digital pins on/off or set a pwm which could be used to control external lights or even motors. Below is the full list of parameters available in environmental.

The returned sensor values from most environmental calls are accessible inside a protocol using Expressions (details). Adjusting protocols in real time based on these values can be very handy for certain types of protocols.

Input: nested array

Example:

"environmental": [[<sensor>], ... ],

Editor: fixed

Instruments:

MultispeQ 1


environmental_array

"environmental_array" defines additional sensors or functionality, just like "environmental". However, "environmental_array" returns sensor data once per pulse, as defined in "pulses". This allows a string of sensor measurements to be used in a single macro. The advantage is you can calculate and return in real time things like rates of change, minima and maxima, etc. See examples for details. Note: "pulsed_lights_brightness" must NOT be set to zero for any given pulse set for the sensors set in "environmental_array" to measure.

Input: nested array

Example:

"environmental_array": [[<number>], ... ],

Editor: fixed

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_length,+ pulse_distance,+ pulsed_lights,+ pulsed_lights_brightness

ir_baseline

Instruments:

MultispeQ 1


label

Each protocol or sub-protocol can now output a label that can be used to indicate something about the protocol, conditions etc.

Input: string

Example:

"label": <string>,

Editor: fixed

Instruments:

MultispeQ 1


max_hold_time

This parameter sets the time (in ms) at which the hold commands timeout. The default value is 15000, or 15 seconds. This value applies to the following control points: start_on_open, start_on_close, start_on_open_close.

Input: number

Example:

"max_hold_time": <number>,

Editor: fixed

Instruments:

MultispeQ 1


message

"message" sends text to the user between pulse sets, and waits for a response before proceeding. Set to "0" to not send a message. "message" should have the same length as "pulses".

Input: nested array

Values:

  • alert - message with an OK button + prompt - message with an OK and Cancel button + confirm - message with an OK and Cancel button and text input box

Example:

"message": [[<message type>, <message>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_length,+ pulse_distance

nonpulsed_lights

"nonpulsed_lights" defines which lights are not pulsed (always on) during a pulse set, and in what order.

Input: nested array

Values:

  • 0 - 10

Example:

"nonpulsed_lights": [[<number>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_distance,+ pulse_length,+ nonpulsed_lights_brightness

nonpulsed_lights_brightness

"nonpulsed_lights_brightness" sets the brightness in microEinsteins (uE) of the nonpulsed lights. Definition of microEinsteins (uE) here.

Input: nested array

Values:

  • 0 - 15000

Example:

"nonpulsed_lights_brightness": [[<number>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_distance,+ pulse_length,+ nonpulsed_lights

number_samples

"number_samples" sets the number of samples taken by the ADC (analog to digital converter) on the sample and hold circuit. The median value of these samples is then saved as a single detector value in "data_raw". The ADC is used to measure the detectors (1 - 4). This value is already set to the optimum value, and most signal noise comes from other source, but in certain cases increasing the value may yield improved signal.

Input: number

Values:

  • 1 - 500

Example:

"number_samples": <number>,

Editor: fixed

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_length,+ pulse_distance,+ detectors

open_close_start

When "open_close_start" is set to 1, the protocol waits until the user fully opens the clamp and closes the clamp before proceeding. Also works if the clamp starts fully open and then closes. Open and close is detected using the Hall effect sensor on the main body and magnet on the clamp (same sensor used to determine sample thickness). If the device is not calibrated, or calibration is off, or the clamping mechanism does not fully open or close, this function will not work. Calibration details can be found by typing print_memory into the console - see thickness_a, thickness_b, thickness_c, thickness_min, and thickness_max for relevant calibration values.

Input: number

Values:

  • 0, 1

Example:

"open_close_start": <number>,

Editor: fixed

Instruments:

MultispeQ 1


par_led_start_on_close

The Ambient light is recreated inside the instrument on closing the clamp. During the hold time, the ambient light is resampled in ~100 ms intervals and the LED output is adjusted to match. When the hold event occurs (e.g. the clamp is closed), the final ambient intensity and LED setting are held and used in the protocol. The number for the LED is defining the LED used for matching the ambient light intensity.

Input: number

Example:

"par_led_start_on_close": <number for LED>,

Editor: fixed

Instruments:

MultispeQ 1


par_led_start_on_open

The Ambient light is recreated inside the instrument on opening the clamp. During the hold time, the ambient light is resampled in ~100 ms intervals and the LED output is adjusted to match. When the hold event occurs (e.g. the clamp is closed), the final ambient intensity and LED setting are held and used in the protocol. The number for the LED is defining the LED used for matching the ambient light intensity.

Input: number

Example:

"par_led_start_on_open": <number for LED>,

Editor: fixed

Instruments:

MultispeQ 1


par_led_start_on_open_close

The Ambient light is recreated inside the instrument on opening the clamp. During the hold time, the ambient light is resampled in ~100 ms intervals and the LED output is adjusted to match. When the hold event occurs (e.g. the clamp is closed), the final ambient intensity and LED setting are held and used in the protocol. The number for the LED is defining the LED used for matching the ambient light intensity.

Input: number

Example:

"par_led_start_on_open_close": <number for LED>,

Editor: fixed

Instruments:

MultispeQ 1


protocols

"protocols" sets the number of times to repeat the protocol. Unlike "averages", this outputs data every time the protocol is repeated (instead of averaging the repeats and outputting data only once).

Input: number

Values:

  • 0 - 999999999

Example:

"protocols": <number>,

Editor: fixed

Instruments:

MultispeQ 1


protocols_delay

"protocols_delay" defines the delay between protocol repeats in in milliseconds (us).

Input: auto

Values:

  • 0 - 9999999999

Example:

"protocols_delay": <number>,

Editor: fixed

Instruments:

MultispeQ 1

Dependancies:

  • protocols

pulse_distance

"pulse_distance" defines the distance, in microseconds (us) between pulses.

Input: array

Values:

  • 750 - 999999999999

Example:

"pulse_distance": [<number>, ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_length,+ pulsed_lights,+ pulsed_lights_brightness

pulse_length

"pulse_length" defines the length, in microseconds (us), of a pulse.

Input: nested array

Values:

  • 1 - 150

Example:

"pulse_length": [[<number>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_distance,+ pulsed_lights,+ pulsed_lights_brightness

pulsed_lights

"pulsed_lights" defines which lights are pulsed, and in what order. When set to 0, no lights are pulsed and no readings are recorded (see Example section here for details).

Input: nested array

Values:

  • 0 - 10

Example:

"pulsed_lights": [[<number>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_distance,+ pulse_length,+ pulsed_lights_brightness

pulsed_lights_brightness

"pulsed_lights_brightness" sets the brightness in microEinsteins (uE) of the pulsed lights. Definition of microEinsteins (uE) here.

Input: nested array

Values:

  • 0 - 15000

Example:

"pulsed_lights_brightness": [[<number>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_distance,+ pulse_length,+ pulsed_lights

pulses

"pulses" defines the number of pulse sets, and quantity of pulses per set.

Input: array

Values:

  • 1 - 8000

Example:

"pulses": [<number>, ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulse_distance,+ pulse_length,+ pulsed_lights,+ pulsed_lights_brightness

recall

This command returns values from the device memory (EEPROM). This includes values saved by the user (see Examples below for details) as well as values saved during factory calibration. For a full list of saved values which can be recalled, see expr.cpp . Expressions may be applied to "recall" requests (more).

Input: array

Example:

"recall": ["userdef[<location>]", ... ],

Editor: fixed

Instruments:

MultispeQ 1


reference

The MultispeQ is designed to be able to nearly simultaneously measure two detectors (within ~4us), normalize and subtract the two signals from each other. This can be useful if there is a optical, electronic (interference), or LED (heating) artifact that needs to be removed. "reference" specifies another detector to be measured and subtracted from "detectors".

Note: It is possible to add two additional detector circuits to the device, on the main body and clamp circuit boards. The devices come unpopulated by default. If added, these two additional detectors are available by settings detectors or reference to 2 (main body) or 4 (clamp). These can then also be used as reference or main detectors.

Input: nested array

Values:

  • 1 - 4

Example:

"reference": [[<detector>], ... ],

Editor: table

Instruments:

MultispeQ 1

Dependancies:

  • pulses,+ pulse_length,+ pulse_distance

save

This command saves a value to a location in the device memory (EEPROM). Expressions can be used on the saved values (more).

Input: nested array

Example:

"save":[[<location, <value>], ... ]],

Editor: fixed

Instruments:

MultispeQ 1


save_trace_time_scale

Saving Time Scale Values. Setting the value to 0 will inactivate the function (preset), setting it to 1 will activate the function. If activated data_raw_time will be added to the measurement, containing the timing information.

Input: number

Values:

  • 0 + 1

Example:

"save_trace_time_scale": <number>,

Editor: fixed

Instruments:

MultispeQ 1


set_led_delay

Pre-illuminating a sample. It is often useful to pre-illuminate a sample at a given light intensity, for a given amount of time.
The following example gives two, 20 second pre-illuminations with the red LED (#2), the first at 0 uE and the second at 100 uE:

"set_led_delay": [
    [2, 20000,0],
    [2, 20000,100]
],

Input: nested array

Example:

"set_led_delay": [ [<number LED>, <number duration in ms>, <number PAR>],... ],

Editor: fixed

Instruments:

MultispeQ 1


set_light_intensity

Instead of measuring the ambient light intensity, use set_light_intensity to set a fixed light intensity. Use the parameters light_intensity or previous_light_intensity to use the defined light intensity.

Note: Instead of defining the light intensity for each pulse set use set_light_intensity to only have one place to change the light intensity.

Input: number

Example:

"set_light_intensity": <number>,

Editor: fixed

Instruments:

MultispeQ 1


start_on_close

If set to 0, the command will be ignored. If set to 1, will wait until the clamp is closed, then proceed with the rest of the experiment.

Input: number

Values:

  • 0 + 1

Example:

"start_on_close": <number>,

Editor: fixed

Instruments:

MultispeQ 1


start_on_open

If set to 0, the command will be ignored. If set to 1 or higher, it will wait until the clamp is opened, then proceed with the rest of the experiment.

Input: number

Values:

  • 0 + 1

Example:

"start_on_open": <number>,

Editor: fixed

Instruments:

MultispeQ 1


start_on_open_close

If set to 0, the command will be ignored. If set to 1, will wait until the clamp is opened then closed, then proceed with the rest of the experiment.

Alias:

  • open_close_start

Input:

Example:

"start_on_open_close": <number>,

Editor: fixed

Instruments:

MultispeQ 1